SL61T / SL61T-A


The data logger will limit the range from -21°C to +71°C. Any temperatures outside this range will be limited to -21°C or +71°C.

 

No damage will occur when using from -41°C to +125°C. Permanent damage will be caused if operated above +125°C. There is a risk of explosion if the data logger is used above +150°C.

 

SL62T / SL62T-A


The data logger limits the range from -41°C to +86°C. Any temperatures outside of this range will be limited to -41°C or +86°C 

 

No damage will occur when using from -41°C to +125°C. Permanent damage will be caused if operated above +125°C. There is a risk of explosion if the data logger is used above +150°C. 


SL63T / SL63T-A


The data logger limits the range from -41°C to +126°C. Any temperatures outside of this range will be limited to -41°C or +126°C 

 

No damage will occur when using from -41°C to +125°C. Permanent damage will be caused if operated above +125°C. There is a risk of explosion if the data logger is used above +150°C. 


SL64TH / SL64TH-A


The data logger limits the temperature range from -21°C to +71°C

 

The data logger limits the humidity range from -1.0% to 101.35%RH

 

No damage will occur when using from -21°C to +80°C, If used temperatures below -21°C and temperature above +81°C There will be a shift in the humidity calibration. Permanent damage will be caused if operated above +125°C. There is a risk of explosion if the data logger is used above +150°C.